• DocumentCode
    3333487
  • Title

    Initial evaluation of direct 4D parametric reconstruction with human PET data

  • Author

    Yan, Jianhua ; Planeta-Wilson, Beata ; Gallezot, Jean-Dominique ; Carson, Richard E.

  • Author_Institution
    PET center, Yale Univ., New Haven, CT, USA
  • fYear
    2009
  • fDate
    Oct. 24 2009-Nov. 1 2009
  • Firstpage
    2503
  • Lastpage
    2506
  • Abstract
    Previously, we presented a direct EM method for producing kinetic parameter images from list mode PET data, where the time-activity curve for each voxel is described by a one-tissue compartment model (1T). The initial evaluations were performed with simulations, without motion, randoms, or scatter effects included. By extension of our previous frame-based physics correction methods, a practical direct 4D parametric reconstruction algorithm is now proposed and implemented for human data. Initial evaluations were performed using 3 human subjects with the serotonin transporter tracer [11C]AFM. Comparisons with the 2-step approach (frame-based reconstruction followed by voxel-by-voxel parameter estimation) provided encouraging initial results. Regional analysis showed that the 2-step and 4D methods have similar K1 and VT values, but with a consistent difference. Visual analysis showed some noise reduction in 4D. These initial results suggest that direct 4D parametric reconstruction can be performed with real data, and offers the potential for improved accuracy and precision over the 2-step frame method.
  • Keywords
    image denoising; image reconstruction; medical image processing; positron emission tomography; [11C]AFM; direct 4D parametric reconstruction; frame-based reconstruction; human PET data; noise reduction; one-tissue compartment model; serotonin transporter tracer; time-activity curve; voxel-by-voxel parameter estimation; Humans; Image reconstruction; Kinetic theory; Noise reduction; Parameter estimation; Performance evaluation; Physics; Positron emission tomography; Reconstruction algorithms; Scattering;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record (NSS/MIC), 2009 IEEE
  • Conference_Location
    Orlando, FL
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-3961-4
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2009.5402049
  • Filename