DocumentCode :
3335103
Title :
Ultra-fast hybrid CPU-GPU Monte Carlo simulation for scatter correction in 3D PETs
Author :
Kim, Kyung Sang ; Ye, Jong Chul
Author_Institution :
Dept. of Bio & Brain Eng., Korea Adv. Inst. of Sci. & Technol., Daejeon, South Korea
fYear :
2011
fDate :
23-29 Oct. 2011
Firstpage :
2749
Lastpage :
2752
Abstract :
A scatter correction algorithm is very important in 3D PET reconstruction due to large scatter contribution in measurements. Currently, one of the most popular methods is single scatter simulation (SSS), which enables a fast calculation of scattering with a relatively good accuracy; however, the accuracy of SSS is dependent on the accuracy of tail fitting to find a correct scaling factor. To overcome this drawback and to improve accuracy of scatter estimation by incorporating multiple scattering contribution, this paper implements a fast Monte Carlo (MC) simulation that considers accurate photon migration and interactions due to photoelectric absorption and Compton scattering. MC simulator provides a prompt and a scatter data separately; hence, a scaling factor can be calculated by comparing a simulated prompt data with a measured data rather than using a tail fitting, which enables a more robust calculation of the scaling factor. Furthermore, we exploit a hybrid CPU-GPU algorithm using the open multi-processing (OpenMP) and CUDA, which results in 124.3 times faster than using a single CPU.
Keywords :
Compton effect; Monte Carlo methods; graphics processing units; hybrid simulation; image reconstruction; medical image processing; multiprocessing systems; parallel architectures; positron emission tomography; 3D PET reconstruction; CUDA; Compton scattering; OpenMP; open multiprocessing; photoelectric absorption; photon migration; scatter correction algorithm; scatter estimation; single scatter simulation; ultrafast hybrid CPU-GPU Monte Carlo simulation; Estimation; Graphics processing unit; Positron emission tomography;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE
Conference_Location :
Valencia
ISSN :
1082-3654
Print_ISBN :
978-1-4673-0118-3
Type :
conf
DOI :
10.1109/NSSMIC.2011.6152961
Filename :
6152961
Link To Document :
بازگشت