DocumentCode :
333599
Title :
In-vitro size estimation of protein clots generated by brain electrodes
Author :
Eriksson, O. ; Wårdell, K.
Author_Institution :
Dept. of Biomed. Eng., Linkoping Univ., Sweden
Volume :
4
fYear :
1998
fDate :
29 Oct-1 Nov 1998
Firstpage :
1783
Abstract :
A method for in-vitro size estimation of protein clots generated by brain electrodes is presented. Radiofrequency generated thermal brain lesions are widely used in functional neurosurgery and in-vitro tests are used to confirm the electrodes´ ability to generate lesions. To be able to estimate the size of protein clots generated in-vitro by brain electrodes, a computer-assisted video system was set up. The size estimation is carried out by software using two captured images of the protein clot. The “true” length and width (9.5 mm) of a sphere as measured with a slide-caliper differed at the most 0.5 mm (5%) and 0.3 mm (3%) respectively, all random errors fall within 2s.d
Keywords :
biomedical electrodes; biomedical measurement; brain; medical image processing; proteins; size measurement; surgery; 9.5 mm; computer-assisted video system; functional neurosurgery; in-vitro size estimation; in-vitro tests; lesions generation; protein clots generated by brain electrodes; radiofrequency generated thermal brain lesions; random errors; slide-caliper measurements; Biomedical electrodes; In vitro; Instruments; Lesions; Neurosurgery; Protein engineering; Radio frequency; Shape; Testing; USA Councils;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 1998. Proceedings of the 20th Annual International Conference of the IEEE
Conference_Location :
Hong Kong
ISSN :
1094-687X
Print_ISBN :
0-7803-5164-9
Type :
conf
DOI :
10.1109/IEMBS.1998.746933
Filename :
746933
Link To Document :