DocumentCode
3336449
Title
Dual detector system for measuring out-of-field neutron dose in proton therapy
Author
Ziebell, A.L. ; Dowdell, S.J. ; Reinhard, M.I. ; Clasie, B. ; Wroe, A. ; Schulte, R.W. ; Lerch, M.L. ; Petaseca, M. ; Perevertaylo, V. ; Anokhin, I.E. ; Rosenfeld, A.B.
Author_Institution
Centre for Med. Radiat. Phys., Univ. of Wollongong, Wollongong, NSW, Australia
fYear
2009
fDate
Oct. 24 2009-Nov. 1 2009
Firstpage
2358
Lastpage
2361
Abstract
A novel technique for separating the out-of-field proton and neutron dose is presented. The technique combines measurements from special pin diodes sensitive to the non ionizing energy losses of protons and neutrons in silicon and standard clinical absorbed dose measurements. Preliminary testing shows good agreement with results from a GEANT4 study.
Keywords
dosimetry; energy loss of particles; p-i-n diodes; radiation therapy; GEANT4 code; dual detector system; nonionizing energy loss; out-of-field neutron dose measurement; pin diodes; proton therapy; Detectors; Energy loss; Energy measurement; Loss measurement; Measurement standards; Medical treatment; Neutrons; Protons; Silicon; Testing;