• DocumentCode
    3336584
  • Title

    Comparison of SOI Microdosimeter and Tissue Equivalent Proportional Counter measurements at the CERF facility

  • Author

    Prokopovich, Dale A. ; Reinhard, Mark I. ; Taylor, Graeme C. ; Hands, Alex ; Rosenfeld, Anatoly B.

  • Author_Institution
    Australian Nucl. Sci. & Technol. Organ. (ANSTO), Lucas Heights, NSW, Australia
  • fYear
    2009
  • fDate
    Oct. 24 2009-Nov. 1 2009
  • Firstpage
    2351
  • Lastpage
    2354
  • Abstract
    The CERN-EU High Energy Reference Field (CERF) facility is used in the calibration of neutron dosimeters for aviation and high energy physics applications. A comparison of the facility´s microdosimetric spectra obtained with a new Silicon on Insulator (SOI) Microdosimeter and a HAWK Tissue Equivalent Proportional Counter (TEPC) are presented. Experimental measurements with a silicon PIN dosimeter diode are also described which indicate a small but significant contribution from charged particles to dosimetry measurements within the neutron dominated field.
  • Keywords
    dosimeters; dosimetry; proportional counters; semiconductor counters; CERF facility; HAWK; SOI microdosimeter; dosimetry measurements; microdosimetric spectra; neutron dosimeters; silicon PIN dosimeter diode; tissue equivalent proportional counter measurements; Australia; Counting circuits; Current measurement; Diodes; Energy measurement; Neutrons; Particle measurements; Performance evaluation; Radiation detectors; Silicon on insulator technology;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record (NSS/MIC), 2009 IEEE
  • Conference_Location
    Orlando, FL
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-3961-4
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2009.5402229
  • Filename
    5402229