DocumentCode
3338307
Title
Development of a model of an X-ray tube transmission source
Author
Goda, Joetta M. ; Ianakiev, Kiril D. ; Moss, Cal E.
Author_Institution
Adv. Technol. Group, Los Alamos Nat. Lab., Los Alamos, NM, USA
fYear
2009
fDate
Oct. 24 2009-Nov. 1 2009
Firstpage
1396
Lastpage
1398
Abstract
In support of the development of an x-ray tube based source for transmission measurements of UF6 gas, we have developed a one-dimensional, spreadsheet-based model of the source. Starting with the spectrum produced by an x-ray tube we apply the linear attenuation coefficients for various notch filters, the aluminum pipe, and UF6 gas. This model allows calculation of the transmitted spectrum based on the type of filter, the thickness of the filter, the x-ray tube high voltage, the Al pipe thickness, and the UF6 gas pressure. The sensitivity of the magnitude of the transmission peak produced by the notch filter to any of these variables can be explored quickly and easily to narrow the choices for experimental measurements. To validate the spreadsheet based model, comparisons have been made to various experimental data.
Keywords
X-ray tubes; notch filters; sensitivity; UF6 gas measurement; X-ray tube; linear attenuation coefficients; magnitude sensitivity; notch filters; one-dimensional spreadsheet-based model; transmission source; transmitted spectrum; Aluminum; Attenuation; CMOS technology; Containers; Laboratories; Nonlinear filters; Nuclear and plasma sciences; Nuclear measurements; Telephony; Voltage;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record (NSS/MIC), 2009 IEEE
Conference_Location
Orlando, FL
ISSN
1095-7863
Print_ISBN
978-1-4244-3961-4
Electronic_ISBN
1095-7863
Type
conf
DOI
10.1109/NSSMIC.2009.5402327