• DocumentCode
    3341429
  • Title

    Development of a 64-channel front-end ASIC for accurate and robust crystal identification of high-resolution PET detectors

  • Author

    Furumiya, Tetsuo ; Ohi, Junichi ; Kitamura, Keishi

  • Author_Institution
    Shimadzu Corp., Kyoto, Japan
  • fYear
    2011
  • fDate
    23-29 Oct. 2011
  • Firstpage
    3901
  • Lastpage
    3903
  • Abstract
    We have developed a 64-channel analog front-end ASIC (application specific integrated circuit) for reading out signals from PS-PMT (position sensitive photomultiplier tube) based high-resolution PET (positron emission tomography) detectors. The ASIC has been designed for accurate and robust crystal identification with Anger logic even under high-count rate conditions. To achieve the feature, a clamp amplifier was built into each readout channel. The whole circuit consists of the front-end containing 64-readout channels, an Anger-type position calculator, and an energy summing amplifier. The readout channel is comprised of current-mode VGA (variable gain amplifier), I/V amplifier, clamp amplifier, and offset controller. In addition, each channel can perform digitally programmable adjustment of gain and offset. The ASIC has been fabricated in the New Japan Radio 0.8-um 2P3M FD-SOI BiCMOS technology and the chip size is 10-mm square. To reduce the mounting area of the ASIC, 196-pin 13-mm square FC-PFBGA (flip chip plastic fine pitch ball grid array) was employed as package of the ASIC. Using this ASIC, we have developed a front-end circuit board for a high-resolution PET scanner and evaluated basic characteristics with a four-layer DOI (depth of interaction) detector that consists of 4,096 LGSO crystal elements and a 64-channnel PS-PMT. As a result, by using the clamp amplifier, each crystal element on the two-dimensional position histogram was clearly identified even at the count rate of 245 kcps.
  • Keywords
    BiCMOS analogue integrated circuits; amplifiers; application specific integrated circuits; ball grid arrays; biomedical electronics; current-mode circuits; flip-chip devices; integrated circuit packaging; photomultipliers; positron emission tomography; solid scintillation detectors; summing circuits; 2P3M FD-SOI BiCMOS technology; 64-channel analog front-end ASIC; 64-channnel PS-PMT; ASIC package; Anger logic; FC-PFBGA; I-V amplifier; LGSO crystal elements; PS-PMT; application specific integrated circuit; chip size; clamp amplifier; current-mode VGA; energy summing amplifier; flip chip plastic fine pitch ball grid array; four-layer DOI detector; front-end circuit board; high-count rate condition; high-resolution PET detectors; high-resolution PET scanner; offset controller; position sensitive photomultiplier tube; positron emission tomography; readout channel; robust crystal identification; size 0.8 mum; two-dimensional position histogram; variable gain amplifier; Arrays; BiCMOS integrated circuits; Clamps; Crystals; Photomultipliers; Silicon;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE
  • Conference_Location
    Valencia
  • ISSN
    1082-3654
  • Print_ISBN
    978-1-4673-0118-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2011.6153741
  • Filename
    6153741