Title :
Cone beam artifact reduction in circular computed tomography
Author :
Chiang, Beshan ; Nakanishi, Satoru ; Zamyatin, Alexander A. ; Shi, Daxin
Abstract :
Cone beam artifact is a well known problem in circular computed tomography. The artifact can be removed by applying exact reconstruction, for example with circle and line trajectory. However, an additional scan often is not available for reconstruction, nor it is practical to collect. Then, methods that can accurately reconstruct image volume from circular data only are of particular interest. In this work we propose a new approach that combines elements of exact reconstruction and iterative reconstruction (SART). We apply iterative procedure to estimate missing line data; at each iteration line data becomes more accurate and cone beam artifacts are further reduced. Evaluation with numerical disc phantom and clinical head data shows effectiveness of the proposed approach in cone beam artifact reduction.
Keywords :
computerised tomography; data analysis; image reconstruction; iterative methods; medical image processing; phantoms; circle trajectory; circular computed tomography; circular data; clinical head data; cone beam artifact reduction; image reconstruction; iteration line data; iterative reconstruction procedure; line trajectory; missing line data estimation; numerical disc phantom data; Biomedical imaging; Filtering; Image reconstruction; Trajectory;
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE
Conference_Location :
Valencia
Print_ISBN :
978-1-4673-0118-3
DOI :
10.1109/NSSMIC.2011.6153789