    Simultaneous reconstruction of activity and attenuation in Time-of-Flight PET

  • Author

    Rezaei, Ahmadreza ; Nuyts, Johan ; Defrise, Michel ; Bal, Girish ; Michel, Christian ; Conti, Maurizio ; Watson, Charles

  • Author_Institution
    Nucl. Med., K.U. Leuven, Leuven, Belgium
  • fYear
    2011
  • fDate
    23-29 Oct. 2011
  • Firstpage
    2375
  • Lastpage
    2382
  • Abstract
    In positron emission tomography (PET) and single photon emission tomography (SPECT), attenuation correction is necessary for quantitative reconstruction of the tracer distribution. Previously, several attempts have been undertaken to estimate the attenuation coefficients from emission data only. These attempts had limited success, because the problem does not have a unique solution, and severe and persistent “cross-talk” between the estimated activity and attenuation distributions was observed. In this paper, we show that the availability of TOF-information eliminates the cross-talk problem by destroying symmetries in the associated Fisher information matrix. We propose a maximum-a-posteriori reconstruction algorithm for jointly estimating the attenuation and activity distributions from TOF-PET data. The performance of the algorithm is studied with 2D simulations, and further illustrated with phantom experiments and with a patient scan. The estimated attenuation image is robust to noise, and does not suffer from the cross-talk that was observed in non-TOF PET. However, some constraining is still mandatory, because the TOF-data determine the attenuation sinogram only up to a constant offset.
  • Keywords
    image reconstruction; medical image processing; phantoms; positron emission tomography; single photon emission computed tomography; time of flight spectra; Fisher information matrix; SPECT; a-posteriori reconstruction algorithm; attenuation sinogram; image simultaneous reconstruction; phantom; positron emission tomography; single photon emission tomography; time-of-flight PET attenuation; tracer distribution; Biomedical imaging; Image resolution; Positron emission tomography;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE
  • Conference_Location
    Valencia
  • ISSN
    1082-3654
  • Print_ISBN
    978-1-4673-0118-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2011.6153883
  • Filename
    6153883