• DocumentCode
    3354944
  • Title

    Digital pulse-shape analysis for Ge Compton camera

  • Author

    Fukuchi, Tomonori ; Motomura, Shinji ; Takeda, Sin´ichiro ; Hiromura, Makoto ; Fukunaka, Ayako ; Haba, Hiromitsu ; Watanabe, Yasuyoshi ; Enomoto, Shuichi

  • Author_Institution
    Center for Mol. Imaging Sci., RIKEN, Kobe, Japan
  • fYear
    2011
  • fDate
    23-29 Oct. 2011
  • Firstpage
    851
  • Lastpage
    852
  • Abstract
    We have developed a real-time digital pulse-shape analysis system for a Ge Compton camera. We successfully demonstrated the improvement of the intrinsic position resolution using the digital pulse-shape analysis. A test measurement with a 54Mn point source showed that the spatial resolution of the Compton camera with the digital system were improved about 1.5 times. The first imaging of an RI phantom, that has 2D hot spots filled with 64Cu of 3 MBq, demonstrated that few millimeter spatial resolution was achieved with our new system.
  • Keywords
    Compton effect; germanium; germanium radiation detectors; nuclear electronics; pulse shaping; 2D hot spots; 54Mn point source; Ge Compton camera; RI phantom; digital pulse-shape analysis system; intrinsic position resolution;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE
  • Conference_Location
    Valencia
  • ISSN
    1082-3654
  • Print_ISBN
    978-1-4673-0118-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2011.6154553
  • Filename
    6154553