• DocumentCode
    335590
  • Title

    Performance of parallel plate ionization chamber for medical irradiation

  • Author

    Urakabe, E. ; Inoue, M. ; Iwashita, Y. ; Shirai, T. ; Sugimura, T. ; Noda, A. ; Kanazawa, M. ; Torikoshi, M. ; Noda, K. ; Yamada, S. ; Tadokoro, M. ; Nishi, M. ; Fujita, Y.

  • Author_Institution
    Inst. for Chem. Res., Kyoto Univ., Japan
  • Volume
    3
  • fYear
    1997
  • fDate
    12-16 May 1997
  • Firstpage
    3819
  • Abstract
    We have developed parallel plate ionization chambers (PPIC) to measure not only the cumulative intensity but also the time structure of slow-extracted heavy-ion beams from a medical synchrotron. The characteristics of the PPIC with 3 mm and 1 mm gap distances for 760, 200, 111, and 55 Torr air were investigated with C6+ beam (290 MeV/u) at HIMAC. The applied voltage to start the plateau region strongly depends on the beam intensity, and pressure of counter gas. The PPIC can be also used as a useful beam monitor for the time-structure measurement of heavy-ion beams
  • Keywords
    ion accelerators; ion beams; ionisation chambers; particle beam diagnostics; particle beam extraction; radiation therapy; synchrotrons; 111 torr; 200 torr; 290 MeV; 55 torr; 760 torr; C; C6+ beam; HIMAC; beam monitor; cumulative intensity measurement; medical irradiation; medical synchrotron; parallel plate ionization chamber performance; plateau region; slow-extracted heavy-ion beams; time structure; Aluminum; Biomedical monitoring; Electrodes; Ion beams; Ionization chambers; Radiation detectors; Radiation monitoring; Synchrotrons; Time measurement; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Particle Accelerator Conference, 1997. Proceedings of the 1997
  • Conference_Location
    Vancouver, BC
  • Print_ISBN
    0-7803-4376-X
  • Type

    conf

  • DOI
    10.1109/PAC.1997.753426
  • Filename
    753426