DocumentCode :
3356662
Title :
Plastic scintillator dosimetry in radiology applications
Author :
Oliveira, Carmen H. ; Rego, Florbela ; Peralta, Luis ; Abreu, Maria C.
Author_Institution :
Lab. de Instrumentacao e Fis. de Particulas, Lisbon, Portugal
fYear :
2011
fDate :
23-29 Oct. 2011
Firstpage :
1612
Lastpage :
1615
Abstract :
Plastics have attractive characteristics that can be exploit in dosimetry. Nevertheless their application with low-energy photon beam is still limited. We aim to a better understanding of their capabilities at the radiology regime. A study of the characteristics of a polystyrene scintillator dosimeter, coupled to a WLS fiber is made in this work. The energy dependence for this detector is extracted for photon beams between 20 and 100 kVp and additional filtrations of 0.5, 1 and 4 mm of Al.
Keywords :
dosimeters; solid scintillation detectors; WLS fiber; low-energy photon beam; plastic scintillator dosimetry; polystyrene scintillator dosimeter; radiology applications; radiology regime; Filtration; Silicon;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE
Conference_Location :
Valencia
ISSN :
1082-3654
Print_ISBN :
978-1-4673-0118-3
Type :
conf
DOI :
10.1109/NSSMIC.2011.6154644
Filename :
6154644
Link To Document :
بازگشت