• DocumentCode
    335986
  • Title

    Estimation of muscle fatigue by using implantable skeletal muscle stimulator

  • Author

    SATO, Eiichi ; Makino, Hideo ; KURODA, Hideo ; AMAN, Jun

  • Author_Institution
    Fac. of Eng., Niigata Inst. of Technol., Japan
  • Volume
    1
  • fYear
    1997
  • fDate
    30 Oct-2 Nov 1997
  • Firstpage
    381
  • Abstract
    In order to investigate the technique of cardiac assistance using skeletal muscle (cardiomyoplasty), the authors are developing an implantable skeletal muscle stimulator that responds to muscle fatigue. The stimulator obtains Fc (as an index of muscle fatigue) by using impulse response from an acceleration sensor. Fc was obtained in the animal experiment, and was compared with ζ/ω by least-squares fitting method using a personal computer. The ζ/ω is interpreted as a time-variant capacitance in the equivalent circuit model. The authors´ results suggest that the muscle fatigue can be estimated by using Fc
  • Keywords
    biomedical electronics; cardiology; prosthetics; surgery; ζ/ω; acceleration sensor; animal experiment; cardiac assistance technique; cardiomyoplasty; equivalent circuit model; implantable skeletal muscle stimulator; impulse response; least-squares fitting method; muscle fatigue estimation; time-variant capacitance; Acceleration; Animals; Capacitance; Cardiology; Equivalent circuits; Fatigue; Heart; Life estimation; Microcomputers; Muscles;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1997. Proceedings of the 19th Annual International Conference of the IEEE
  • Conference_Location
    Chicago, IL
  • ISSN
    1094-687X
  • Print_ISBN
    0-7803-4262-3
  • Type

    conf

  • DOI
    10.1109/IEMBS.1997.754556
  • Filename