  • Title

    Study on ultrasound transducer with double peak type frequency characteristics for sub-harmonic imaging

  • Author

    Takeuchi, Shinichi ; Al Zaabi, Mohamed Rashed Ali ; Sato, Toshio ; Kawashima, Norimichi

  • Author_Institution
    Fac. of Eng., Toin Univ., Kanagawa, Japan
  • Volume
    2
  • fYear
    2002
  • fDate
    8-11 Oct. 2002
  • Firstpage
    1101
  • Abstract
    In this study, an ultrasound transducer with double peak type frequency characteristics having a peak on the high frequency side for illuminating ultrasound to tissues or microbubbles in the human body and a peak on the low frequency side for detecting sub-harmonic component from microbubbles was designed with numerical calculations and fabricated as the one for sub-harmonic imaging. The trial ultrasound transducer was confirmed to have the double peak type frequency characteristics of transmitting and receiving sensitivity adapted to sub-harmonic imaging.
  • Keywords
    biomedical ultrasonics; ultrasonic imaging; ultrasonic transducers; double peak type frequency characteristics; human body; microbubbles; numerical calculations; receiving sensitivity; sub-harmonic imaging; tissues; transmitting sensitivity; ultrasound transducer; Acoustic transducers; Biomedical imaging; Biomedical transducers; Ceramics; Equivalent circuits; Frequency; Impedance; Piezoelectric transducers; Ultrasonic imaging; Ultrasonic transducers;