• DocumentCode
    336562
  • Title

    Design of a near infrared reflectometer

  • Author

    Pasley, Richard L. ; Wolf, Erich ; Bada, Henrietta ; Leffler, Charles W. ; Daley, Michael L.

  • Author_Institution
    Dept. of Electr. Eng., Memphis Univ., TN, USA
  • Volume
    2
  • fYear
    1997
  • fDate
    30 Oct-2 Nov 1997
  • Firstpage
    729
  • Abstract
    A two channel near infrared reflectometer was designed and fabricated using laser (814 nm and 751 nm) diodes and separate photodiode detector circuits. Because the isosbestic point of HbO2 and Hb is approximately 814 nm, the gains of each channel were equated at this wavelength. The device is designed to measure relative change of local cerebral blood volume, HbO2 and Hb concentration in an unknown sample volume
  • Keywords
    biomedical electronics; blood; brain; laser applications in medicine; photodiodes; reflectometers; semiconductor lasers; volume measurement; 751 nm; 814 nm; Hb concentration; HbO2; O2; hemoglobin; isosbestic point; laser diodes; local cerebral blood volume; medical instrumentation; photodiode detector circuit; reflectometer design; relative change measurement; two channel near infrared reflectometer; unknown sample volume; Blood; Circuits; Diodes; Infrared detectors; Optical design; Particle scattering; Reflectometry; Signal detection; Volume measurement; Wavelength measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1997. Proceedings of the 19th Annual International Conference of the IEEE
  • Conference_Location
    Chicago, IL
  • ISSN
    1094-687X
  • Print_ISBN
    0-7803-4262-3
  • Type

    conf

  • DOI
    10.1109/IEMBS.1997.757739
  • Filename
    757739