SPECT-based radioimmunotherapy planning system
Author :
Laitinen, J. ; Alakuijala, J. ; Helminen, H. ; Sallinen, S.
Author_Institution :
Varian-Dosetek Oy, Espoo
Volume :
2
fYear :
1997
fDate :
30 Oct-2 Nov 1997
Firstpage :
781
Abstract :
A 3D treatment planning system for radioimmunotherapy based on activity maps from SPECT images has been developed. Absorbed dose distributions are calculated using a point source kernel convolution with the activity map using FFT. Patient specific absorbed dose can be calculated and analyzed volumetrically together with diagnostic images. The method overcomes the limitations of conventional radioimmunotherapy dose modelling and provides the means for optimizing treatment strategy for each patient. The system offers a wide range of dose analysis and image processing tools. Connectivity to imaging devices is handled using industry standards
Keywords :
dosimetry; fast Fourier transforms; medical image processing; radiation therapy; single photon emission computed tomography; 3D treatment planning system; SPECT-based radioimmunotherapy planning system; absorbed dose distributions; connectivity to imaging devices; dose modelling; industry standards; medical diagnostic imaging; nuclear medicine; optimized treatment strategy; patient specific absorbed dose; point source kernel convolution; Computed tomography; Convolution; Image analysis; Kernel; Magnetic resonance imaging; Medical treatment; Oncology; Optical imaging; Visualization; X-ray imaging;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 1997. Proceedings of the 19th Annual International Conference of the IEEE
Conference_Location :
Chicago, IL
ISSN :
1094-687X
Print_ISBN :
0-7803-4262-3
Type :
conf
DOI :
10.1109/IEMBS.1997.757755