Title :
Fabrication of a 3D laser imager for measuring wound geometry
Author :
Rogers, Bill ; Bunegin, Leonid ; Tolstykh, Gleb ; Walsh, Nicolas ; Smith, R. Brian
Author_Institution :
Health Sci. Center, Texas Univ., San Antonio, TX, USA
fDate :
30 Oct-2 Nov 1997
Abstract :
Chronic wounds are a large part of daily medical practice, and are an increasing focus of research. As the number of modalities to treat chronic wounds increase, the validity of these modalities must be tested. A low cost laser imager was designed and fabricated for the measurement of wound geometry. The accuracy of the imager was validated using simulated wounds of known geometry. Perimeter, area, and volume measurements were compared with conventional measurement techniques on wound models simulated on plaster mockups of the lower leg and foot
Keywords :
area measurement; biomedical equipment; biomedical imaging; biomedical measurement; laser applications in medicine; skin; volume measurement; 3D laser imager fabrication; chronic wounds; daily medical practice; foot; low cost laser imager; lower leg; medical instrumentation; perimeter measurements; plaster mockups; simulated wounds; wound geometry measurement; wound models; Biomedical imaging; Costs; Focusing; Geometrical optics; Laser modes; Optical design; Optical device fabrication; Solid modeling; Testing; Wounds;
Conference_Titel :
Engineering in Medicine and Biology Society, 1997. Proceedings of the 19th Annual International Conference of the IEEE
Conference_Location :
Chicago, IL
Print_ISBN :
0-7803-4262-3
DOI :
10.1109/IEMBS.1997.757806