DocumentCode
3368194
Title
Ringdown reduction for an intraluminal ultrasound array using depth-dependent azimuthal filters
Author
de Ana, F.J. ; O´Donnell, M.
Author_Institution
Dept. of Biomed. Eng., Michigan Univ., Ann Arbor, MI, USA
Volume
2
fYear
2002
fDate
8-11 Oct. 2002
Firstpage
1773
Abstract
Vibration of piezoelectric elements after electrical excitation creates ringdown artifacts in ultrasound images close to the transducer surface. For intraluminal applications, high quality images are often needed at depths close to the catheter surface. Consequently, ringdown artifacts pose a serious problem for intraluminal circumferential arrays. A current method to reduce these artificial echoes subtracts a precalibrated ringdown signal from each frame in a data set. The dynamic nature of the ringdown artifact, however, reduces the effectiveness of this subtraction algorithm with time following data collection. The ringdown signal has very distinctive characteristics, especially in the near field azimuthal spectrum. An in-depth study of those characteristics presents new alternatives to minimize ringdown artifacts. Using a set of depth-dependent azimuthal filters, we present a method to significantly reduce ringdown artifacts while minimally affecting the real signal. Rubber phantoms with embedded graphite particles were imaged close to the transducer surface to study potential improvements obtained with the azimuthal filters. Results show a reduction of the ringdown signal by at least 10 dB, while increasing the dynamic range between the real signal and the ringdown from 2 dB to 15 dB. The combination of azimuthal filtering with ringdown subtraction can greatly enhance contrast at ranges close to the transducer surface.
Keywords
biomedical transducers; biomedical ultrasonics; piezoelectric transducers; ultrasonic transducer arrays; azimuthal filtering; depth-dependent azimuthal filters; electrical excitation; high quality images; intraluminal applications; intraluminal circumferential arrays; intraluminal ultrasound array; near field azimuthal spectrum; piezoelectric elements vibrations; ringdown reduction; rubber phantoms; subtraction algorithm; time following data collection; transducer surface; ultrasound images; Catheters; Dynamic range; Filtering; Filters; Imaging phantoms; Piezoelectric transducers; Rubber; Ultrasonic imaging; Ultrasonic transducer arrays; Ultrasonic transducers;
fLanguage
English
Publisher
ieee
Conference_Titel
Ultrasonics Symposium, 2002. Proceedings. 2002 IEEE
ISSN
1051-0117
Print_ISBN
0-7803-7582-3
Type
conf
DOI
10.1109/ULTSYM.2002.1192642
Filename
1192642