DocumentCode
3370353
Title
A frequency-shift readout system for FPW allergy biosensor
Author
Hsu, Chia-Hao ; Lin, Yain-Reu ; Tsai, Yue-Da ; Chen, Yun-Chi ; Wang, Chua-Chin
Author_Institution
Dept. of Electr. Eng., Nat. Sun Yat-Sen Univ., Kaohsiung, Taiwan
fYear
2011
fDate
2-4 May 2011
Firstpage
1
Lastpage
4
Abstract
In this paper, an IgE antigen concentration measurement system using a frequency-shift readout method for a two-port FPW (flexural plate-wave) allergy biosensor is presented. The proposed frequency-shift readout method adopts a peak detecting scheme to detect the resonant frequency. A linear frequency generator, a pair of peak detectors, two registers, and an subtractor are only needed in our system. According to the specification of the FPW allergy biosensor, the frequency sweep range is limited in 2 MHz to 10 MHz. The sensitivity of the peak detector is 0.8 mV. The proposed frequency-shift readout circuit is verified on silicon by using a standard 0.18 μm CMOS technology. The maximal power consumption is 12.94 mW@0.1 MHz clock given by HSPICE simulations.
Keywords
CMOS integrated circuits; SPICE; biosensors; readout electronics; CMOS technology; FPW allergy biosensor; HSPICE simulations; IgE antigen concentration measurement system; flexural plate-wave; frequency-shift readout system; Biological system modeling; Biosensors; Detectors; Immune system; Oscillators; Registers; Resonant frequency; FPW; IgE antigen; frequency-shift readout circuit; peak detection; resonant frequency;
fLanguage
English
Publisher
ieee
Conference_Titel
IC Design & Technology (ICICDT), 2011 IEEE International Conference on
Conference_Location
Kaohsiung
ISSN
Pending
Print_ISBN
978-1-4244-9019-6
Type
conf
DOI
10.1109/ICICDT.2011.5783241
Filename
5783241