DocumentCode
3372824
Title
A new method of decoupling for gradient systems in MRI
Author
Guang Chen ; Wenhui Yang ; Xing Lv
Author_Institution
Inst. of Electr. Eng., Beijing, China
fYear
2013
fDate
16-18 Dec. 2013
Firstpage
1
Lastpage
5
Abstract
A set of external coils is designed to eliminate the induced voltage in shim coils caused by the changing current in the gradient coils. By placing the external coil near the line which supplies the current for the gradient coil in a certain way and inserting it into the circuit of shim coils, a induced voltage which has the same wave form with the induced voltage in the shim coil is generated and consequently the coupling between the shim coil and the gradient coil is eliminated. Moreover, the induced voltages of various shim coils are simulated and the parameter of one external coil is determined by the result of the simulation.
Keywords
biomedical MRI; coils; gradient methods; shims; waveform analysis; MRI; decoupling; external coils; gradient coils; gradient systems; magnetic resonance imaging; shim coils; waveform; Coils; Inductance; Magnetic flux; Magnetic resonance imaging; Magnetic separation; Mathematical model; Solid modeling; MRI; decoupling; external coils; simulation;
fLanguage
English
Publisher
ieee
Conference_Titel
Biomedical Engineering and Informatics (BMEI), 2013 6th International Conference on
Conference_Location
Hangzhou
Print_ISBN
978-1-4799-2760-9
Type
conf
DOI
10.1109/BMEI.2013.6746896
Filename
6746896
Link To Document