Title :
Selective stimulation of a peripheral nerve: correction of drop foot in three patients
Author :
Rozman, J. ; Tekavcic, I.
Author_Institution :
Centre for Implantable Technol. & Sensors, Ljubljana, Slovenia
fDate :
31 Oct-3 Nov 1996
Abstract :
We present an evaluation of an implantable system with a monopolar half-cuff for selective stimulation of the common peroneal nerve in three hemiplegic patients. The stimulus proposed in this work for effecting selective stimulation of the superficial region of the human common peroneal nerve in more physiological order were current, charge balanced, and biphasic pulses with a rectangular cathodic component, and exponential decay anodic component delayed for 50 μs. Results show that a half-cuff electrode is capable of making a long term selective activation, mainly of those muscles that contribute to strong dorsal flexion and moderate eversion of the hemiplegic´s foot. Moreover, significant improvements of gait dynamics without excessive eversion were observed
Keywords :
biocontrol; bioelectric phenomena; biomechanics; biomedical telemetry; handicapped aids; muscle; neurophysiology; orthotics; patient treatment; radiotelemetry; 2 MHz; biphasic pulses; charge balanced pulses; common peroneal nerve; current pulses; drop foot correction; exponential decay anodic component; gait dynamics; human common peroneal nerve; implantable system; long term selective activation; monopolar half-cuff; peripheral nerve; rectangular cathodic component; selective stimulation; strong dorsal flexion; superficial region; three hemiplegic patients; Delay effects; Electrodes; Foot; Humans; Leg; Muscles; Nervous system; Recruitment; Sensor systems; Switches;
Conference_Titel :
Engineering in Medicine and Biology Society, 1996. Bridging Disciplines for Biomedicine. Proceedings of the 18th Annual International Conference of the IEEE
Conference_Location :
Amsterdam
Print_ISBN :
0-7803-3811-1
DOI :
10.1109/IEMBS.1996.646522