DocumentCode
3378547
Title
Identifying artifacts in arterial blood pressure using morphogram variability
Author
Ali, W. ; Eshelman, L. ; Saeed, M.
fYear
2004
fDate
19-22 Sept. 2004
Firstpage
697
Lastpage
700
Abstract
In this article we present a simple technique that utilizes the cross correlations between ECG signals and haemodynamic signals for the purpose of assessing signal quality and detecting artifacts in the arterial blood pressure (ABP) signal. The technique was tested using cases from a physician-annotated patient monitoring signal database from Beth IsraeUHarvard-MIT University data bank. The results were encouraging: 90% of the manually annotated artifacts were correctly classified as artifacts and 99% of the manually annotated true events were correctly classified (out of a tQtd Of 683 manualiy annotated alarms).
Keywords
Arterial blood pressure; Biomedical monitoring; Blood flow; Blood pressure; Electrocardiography; Heart beat; Patient monitoring; Signal analysis; Signal detection; Signal processing;
fLanguage
English
Publisher
ieee
Conference_Titel
Computers in Cardiology, 2004
Conference_Location
Chicago, IL, USA
Print_ISBN
0-7803-8927-1
Type
conf
DOI
10.1109/CIC.2004.1443034
Filename
1443034