Abstract
Multisensor monitoring devices that use skin surface or implanted sensors are susceptible to changes in temperature, sweat, and movement, such that the measured data cannot be used. This paper presents an automatic approach to detect such erroneous sensors. It is based on the assumption that valid measurements are related by a reconstruction model, while measurements from erroneous sensors are unrelated. The method estimates the data at each sensor based on the measurements from all other sensors, and compares it to the measurements. The sensor-data match is tested using ANOVA to detect the presence of an erroneous sensor. The method was tested on simulated and experimental data of electrical impedance tomography (EFT) and showed consistent identification of erroneous electrodes
Keywords
biomedical imaging; biomedical measurement; electric impedance imaging; sensor fusion; skin; statistical analysis; tomography; ANOVA; EFT; automatic detection; data errors; electrical impedance tomography; erroneous electrodes; erroneous sensors; implanted sensors; measurement reliability; multisensor monitoring devices; reconstruction model; sensor-data match; skin surface; Biomedical measurements; Biomedical monitoring; Chemical technology; Electrodes; Sensor phenomena and characterization; Sensor systems; Skin; Temperature sensors; Testing; Tomography; Data errors; reconstruction algorithms; sensors;
fLanguage
English
Publisher
ieee
Conference_Titel
Instrumentation and Measurement Technology Conference, 2005. IMTC 2005. Proceedings of the IEEE
Conference_Location
Ottawa, Ont.
Print_ISBN
0-7803-8879-8
Type
conf
DOI
10.1109/IMTC.2005.1604444
Filename
1604444
Link To Document