DocumentCode :
3380098
Title :
New Column Designs for MicroGC
Author :
Radadia, Adarsh D. ; Masel, Richard I. ; Shannon, Mark A.
Author_Institution :
Univ. of Illinois at Urbana-Champaign, Urbana
fYear :
2007
fDate :
10-14 June 2007
Firstpage :
2011
Lastpage :
2014
Abstract :
In this paper, we present some new column designs for micro-gas chromatography (muGC) columns. Serpentine columns are compared to commonly used spiral columns and are found to give relatively lower height equivalent to theoretical plate (HETP) values. Turns are a major dispersion source for serpentine configuration. We found that sinusoidal compensation structures following the turns give lowest dispersion amongst the several tested turn geometries for serpentine configuration.
Keywords :
chromatography; gas sensors; microsensors; column design; micro-gas chromatography column; microGC; serpentine column; serpentine configuration; sinusoidal compensation structure; Actuators; Design engineering; Electrokinetics; Fabrication; Geometry; Microchannel; Silicon; Solid state circuits; Spirals; Testing; dispersion in microchannels; microGC; microcolumns;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Solid-State Sensors, Actuators and Microsystems Conference, 2007. TRANSDUCERS 2007. International
Conference_Location :
Lyon
Print_ISBN :
1-4244-0842-3
Electronic_ISBN :
1-4244-0842-3
Type :
conf
DOI :
10.1109/SENSOR.2007.4300557
Filename :
4300557