DocumentCode
3388118
Title
Performance comparison of low current measurement systems for biomedical applications
Author
Kim, Dongsoo ; Tang, Wei ; Goldstein, Brian ; Weerakoon, Pujitha ; Montanaro, Hazael ; Martini, Berin ; Culurciello, Eugenio
Author_Institution
Dept. of Electr. Eng., Yale Univ., New Haven, CT, USA
fYear
2010
fDate
May 30 2010-June 2 2010
Firstpage
3469
Lastpage
3472
Abstract
In this paper, we report on the noise analysis of low current measurement systems for biomedical applications. We analyzed resistive feedback, capacitive feedback and current conveyor circuits for low current measurement systems. Detailed noise analysis are presented and matched with measurement data using a 0.5-μm fabrication process. Based on the theoretical analysis and the measurements, the capacitive feedback system provides better noise performance for the measurement of low current. The capacitive feedback is capable of measuring 700fA RMS at 10KHz sampling rate, whereas the resistive feedback provides 4pA and the current conveyor provides 600pA. This paper provides design guidelines to maximize the measurement performance of low current for biomedical instrumentation.
Keywords
bioelectric phenomena; feedback; noise; biomedical application; capacitive feedback; current conveyor circuit; low current measurement system; noise performance; performance comparison; resistive feedback; Biomedical measurements; Circuit noise; Current measurement; DNA; Feedback; Immune system; Impedance; Integrated circuit measurements; Noise measurement; Performance analysis;
fLanguage
English
Publisher
ieee
Conference_Titel
Circuits and Systems (ISCAS), Proceedings of 2010 IEEE International Symposium on
Conference_Location
Paris
Print_ISBN
978-1-4244-5308-5
Electronic_ISBN
978-1-4244-5309-2
Type
conf
DOI
10.1109/ISCAS.2010.5537843
Filename
5537843
Link To Document