• DocumentCode
    340177
  • Title

    A new γ-probe for radioassisted oncological surgery

  • Author

    Dusi, W. ; Bollini, D. ; Moroni, C. ; Ricard, M.

  • Author_Institution
    Ist. TESRE, CNR, Bologna, Italy
  • Volume
    2
  • fYear
    1998
  • fDate
    1998
  • Firstpage
    754
  • Abstract
    The operating principle and preliminary results of a new hand-held non imaging γ-probe are presented and its use in radioassisted oncological surgery is discussed. The probe, having an external diameter of about 17 mm, is based on a square array of four 5×5 mm2 CdTe semiconductor detectors read separately by each own electronic circuit. An original handling of the signals allows the probe to provide at the same time both the total counting rate as well as an indication on the value of displacement of the probe axis from the center of the emitting source and the direction towards which move to center. Using this probe during surgical operation a significant improvement in a faster and more accurate localization of gamma-emitting regions will be possible with respect to those obtained with a conventional single detector probe
  • Keywords
    gamma-ray apparatus; gamma-ray detection; radioisotope imaging; semiconductor counters; surgery; tumours; 17 mm; 5 mm; CdTe; hand-held nonimaging γ-probe; radioassisted oncological surgery; semiconductor detectors; Cancer; Gamma ray detection; Gamma ray detectors; Malignant tumors; Oncological surgery; Pixel; Probes; Sensor arrays; Solid scintillation detectors; Spatial resolution;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium, 1998. Conference Record. 1998 IEEE
  • Conference_Location
    Toronto, Ont.
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-5021-9
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1998.774284
  • Filename
    774284