Title :
DC rejection and deblocking in multichannel bioelectric recordings
Author :
Grimbergen, C.A. ; MettingVanRijn, A.C. ; Kuiper, A.P. ; Honsbeek, R.H. ; Speijer, K. ; Peper, A.
Author_Institution :
Dept. of Med. Phys., Amsterdam Univ., Netherlands
Abstract :
Multichannel instrumentation for the recording of bioelectric events should have provisions to reject the relatively large DC voltages which are produced by the electrodes involved. Active DC rejection is a suitable high pass filtering technique preserving the common mode rejection ratio of the preamplifiers. High pass filtering may lead to very slow responses of the instrumentation after overload, which can be handled by deblocking. A method to use the DC information of the rejected signal in a multichannel system is presented
Keywords :
adaptive signal processing; bioelectric potentials; biomedical electronics; high-pass filters; medical signal processing; DC rejection; active DC rejection; bioelectric events recording; common mode rejection ratio; deblocking; high pass filtering; high pass filtering technique; multichannel bioelectric recordings; multichannel instrumentation; multichannel system; overload; preamplifiers; Bioelectric phenomena; Biomedical signal processing; Circuits; Electrodes; Filtering; Filters; Frequency; Hardware; Signal processing; Signal to noise ratio;
Conference_Titel :
Engineering in Medicine and Biology Society, 1995., IEEE 17th Annual Conference
Conference_Location :
Montreal, Que.
Print_ISBN :
0-7803-2475-7
DOI :
10.1109/IEMBS.1995.579881