DocumentCode :
3404487
Title :
Calibration of measurement system based on phase measurement profilometry
Author :
Zuo-chun, Shen ; Hua-wei, Li ; Yan-zhou, Zhou
Author_Institution :
Sch. of Astronaut., Harbin Inst. of Technol. Harbin, Harbin, China
fYear :
2011
fDate :
12-16 Oct. 2011
Firstpage :
200
Lastpage :
203
Abstract :
With the phase measurement profilometry widely used, the precision phase measurement profilometry is demanded. The system was calibrated using by translating the calibration plate. Relationship between the coordinate z and phase was determined. And the coordinates (x, y, z) and pixel coordinates (m, n) relationships were determined too. We analyzed the results of each step of the standard deviation of calibration and get a high calibration accuracy. In order to further improve the calibration accuracy, we corrected the angle errors between the camera coordinate system and the actual world coordinate system. In order to verify the validity of this calibration method, we measured a ball bearing. The measurement accuracy is 0.0428+/-0.01mm.
Keywords :
calibration; measurement errors; surface topography measurement; calibration accuracy; measurement accuracy; measurement system; phase measurement; profilometry; Accuracy; Ball bearings; Calibration; Cameras; Coordinate measuring machines; Fitting; Phase measurement; Angle errors; Phase measurement profilometry; Standard deviation; System calibration;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Optoelectronics and Microelectronics Technology (AISOMT), 2011 Academic International Symposium on
Conference_Location :
Harbin
Print_ISBN :
978-1-4577-0794-0
Type :
conf
DOI :
10.1109/AISMOT.2011.6159353