DocumentCode
340695
Title
Correction for recombination loss of ions in ionization chambers
Author
Takata, N. ; Yin, Zaizhe ; Li, Aiwu
Author_Institution
Electrotech. Lab., Ibaraki, Japan
Volume
1
fYear
1998
fDate
1998
Firstpage
558
Abstract
The signal charge loss due to initial recombination and that due to volume recombination have been measured separately using parallel plate ionization chambers irradiated with cobalt-60 gamma rays, over a wide range of applied voltage and air humidity. It is shown that both initial and volume recombination depend on the electric field strength and also on humidity through ion clustering reaction with water molecules. In spite of the usual assumptions, a linear relation has not been observed between the inverses of the signal current and applied voltage for an initial recombination, nor between the inverses of the signal current and square of the applied voltage for volume recombination. Methods and data which can be used for estimation of initial and volume recombination losses of ions in ionization chambers are proposed
Keywords
gamma-ray effects; ion recombination; ionisation chambers; H2O; applied voltage; electric field strength; gamma rays; initial recombination; ionization chambers; ions; parallel plate ionization chambers; recombination loss; signal current; volume recombination; water molecules; Electric potential; Electrodes; Equations; Humidity; Ionization chambers; Laboratories; Metrology; Spontaneous emission; Voltage; Volume measurement;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium, 1998. Conference Record. 1998 IEEE
Conference_Location
Toronto, Ont.
ISSN
1082-3654
Print_ISBN
0-7803-5021-9
Type
conf
DOI
10.1109/NSSMIC.1998.775203
Filename
775203
Link To Document