• DocumentCode
    340695
  • Title

    Correction for recombination loss of ions in ionization chambers

  • Author

    Takata, N. ; Yin, Zaizhe ; Li, Aiwu

  • Author_Institution
    Electrotech. Lab., Ibaraki, Japan
  • Volume
    1
  • fYear
    1998
  • fDate
    1998
  • Firstpage
    558
  • Abstract
    The signal charge loss due to initial recombination and that due to volume recombination have been measured separately using parallel plate ionization chambers irradiated with cobalt-60 gamma rays, over a wide range of applied voltage and air humidity. It is shown that both initial and volume recombination depend on the electric field strength and also on humidity through ion clustering reaction with water molecules. In spite of the usual assumptions, a linear relation has not been observed between the inverses of the signal current and applied voltage for an initial recombination, nor between the inverses of the signal current and square of the applied voltage for volume recombination. Methods and data which can be used for estimation of initial and volume recombination losses of ions in ionization chambers are proposed
  • Keywords
    gamma-ray effects; ion recombination; ionisation chambers; H2O; applied voltage; electric field strength; gamma rays; initial recombination; ionization chambers; ions; parallel plate ionization chambers; recombination loss; signal current; volume recombination; water molecules; Electric potential; Electrodes; Equations; Humidity; Ionization chambers; Laboratories; Metrology; Spontaneous emission; Voltage; Volume measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium, 1998. Conference Record. 1998 IEEE
  • Conference_Location
    Toronto, Ont.
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-5021-9
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1998.775203
  • Filename
    775203