Title :
Performance of a gas proportional scintillation counter with compact modular high voltage and optocoupled electronics
Author :
Silva, Rui M C ; Conde, C.A.N.
Author_Institution :
Dept. de Fisica, Coimbra Univ., Portugal
Abstract :
Improved compact (less than 2 dm3), low weight (less than 1 kg) and low power (less than 5 W) electronic circuits for large window area (20 cm2) Gas Proportional Scintillation Counters (GPSCs) with an ellipsoidal grid have been developed for the detection of solar X-rays in balloon born experiments. The electronics consist of amplification stages, modular power supplies, including a 6 kV detector supply and a circuit for optically coupling the detector pulses to ground. X-ray spectra demonstrating the performance of the GPSC with the compact electronics are presented for 109Cd and 241Am X-rays. The system energy resolution for 22.1 keV is 4.7% and for 59.6 keV is 3.7%
Keywords :
X-ray detection; cosmic ray apparatus; gas scintillation detectors; nuclear electronics; opto-isolators; proportional counters; solar radiation; 6 kV; 109Cd; 241Am; X-ray spectra; amplification stages; balloon born experiments; compact modular high voltage; gas proportional scintillation counter; gas proportional scintillation counters; modular power supplies; optocoupled electronics; solar X-rays; system energy resolution; Coupling circuits; Electronic circuits; Optical coupling; Optical detectors; Pulsed power supplies; Scintillation counters; Solid scintillation detectors; Stimulated emission; X-ray detection; X-ray detectors;
Conference_Titel :
Nuclear Science Symposium, 1998. Conference Record. 1998 IEEE
Conference_Location :
Toronto, Ont.
Print_ISBN :
0-7803-5021-9
DOI :
10.1109/NSSMIC.1998.775236