DocumentCode :
3410384
Title :
Rapid data acquisition procedures for cardiac MRI
Author :
Kelley, D.A.C. ; Kramer, D.M. ; Crooks, L.E. ; Kaufman, L.
Author_Institution :
California Univ., San Francisco, CA, USA
fYear :
1988
fDate :
4-7 Nov. 1988
Firstpage :
325
Abstract :
The authors have developed techniques which may be performed on any standard clinical imager and produce complete data sets of 16 time frames in each of eight contiguous slices in a total acquisition time of 1024 heartbeats. The limits inherent in the acquisition procedure are outlined. The authors have found that 3.0 mm in-plane resolution is sufficient, and consequently, they need only perform 64 phase-encoding steps in a transaxial orientation and 128 in a coronal to produce a sufficiently large field of view to avoid aliasing in all but large patients. They have implemented these techniques using synchronous gating in both a spin echo (RF refocused) and a gradient reversal single pulse (gradient refocused) acquisition procedure. The former will suppress signal from flowing blood, thus producing a clean image of the cardiac anatomy, and that the latter will produce an image dominated by the inflowing blood due to flow related enhancement, possibly revealing variations in the flow due to abnormalities.<>
Keywords :
biomedical NMR; blood; cardiology; data acquisition; haemodynamics; medical computing; RF refocused pulse; acquisition time; cardiac MRI; cardiac anatomy; clinical imager; data acquisition; flowing blood; gradient reversal single pulse; heartbeats; phase-encoding steps; spin echo pulse; synchronous gating;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 1988. Proceedings of the Annual International Conference of the IEEE
Conference_Location :
New Orleans, LA, USA
Print_ISBN :
0-7803-0785-2
Type :
conf
DOI :
10.1109/IEMBS.1988.94541
Filename :
94541
Link To Document :
بازگشت