Title :
Medical Equipment Immunity Assessment by Time Domain Analysis Application to GSM signals
Author :
Fernandez-Chimeno, M. ; García-González, Miguel Ángel ; Silva, Ferran
Author_Institution :
Tech. Univ. of Catalonia, Barcelona
Abstract :
This paper presents a measurement method to assess the immunity of medical equipment (bioelectrical signal recording systems) to short duration interferences, like that of a GSM signal during the call establishment and ringing phase. Frequency domain measurement methods do not allow the analysis of these signals due to their short duration and transient nature. To measure these interferences we use the intermediate frequency (IF) signal output of a spectrum analyzer working in "zero span" to obtain a signal in time domain, and we connect it to the input of a digital oscilloscope. Then, we demodulate the IF signal to obtain the interference in base band. The combination of the characteristics of both instruments, allow us to measure the interfering signal, and to assess the effect of these kinds of signals in medical systems that record low-level, low-frequency signals
Keywords :
bioelectric potentials; biomedical equipment; cellular radio; immunity testing; radiofrequency interference; time-domain analysis; GSM signals; bioelectrical signal recording systems; digital oscilloscope; intermediate frequency signal output; medical equipment immunity assessment; short duration interferences; spectrum analyzer; time domain analysis; zero span; Bioelectric phenomena; Biomedical equipment; Frequency domain analysis; Frequency measurement; GSM; Immune system; Interference; Phase measurement; Signal analysis; Time domain analysis;
Conference_Titel :
Electromagnetic Compatibility, 2007. EMC 2007. IEEE International Symposium on
Conference_Location :
Honolulu, HI
Print_ISBN :
1-4244-1349-4
Electronic_ISBN :
1-4244-1350-8
DOI :
10.1109/ISEMC.2007.62