Title :
Microwave-induced thermoelastic tissue imaging
Author :
Chan, Karen H. ; Lin, James C.
Author_Institution :
Dept. of Bioeng., Illinois Univ., Chicago, IL, USA
Abstract :
A prototype microwave-induced thermoelastic tissue imaging system was designed to acquire and process two-dimensional projections of various sizes test tubes filled with tissue equivalent materials. A test object is immersed in a tank of water at whose surface single microwave pulses of 2 mu s at a carrier frequency of 2450 MHz are launched. Thermoelastic waves induced at the water surface are detected, on propagating through the object material, by a 20*20 piezoelectric transducer array at the bottom of the water tank. A computer-controlled data acquisition system samples and converts the amplified and bandlimited signals into digital form. The system produces images interactively through the use of image processing techniques. Results are presented to demonstrate the potential use of microwave-induced thermoelastic waves as a possible medical imaging modality.<>
Keywords :
microwave imaging; patient diagnosis; thermoelasticity; 2450 MHz; 2D projections; amplified bandlimited signals; computer-controlled data acquisition system; image processing; medical diagnostic imaging; microwave-induced thermoelastic tissue imaging; piezoelectric transducer array; test tubes; tissue equivalent materials; water surface; water tank;
Conference_Titel :
Engineering in Medicine and Biology Society, 1988. Proceedings of the Annual International Conference of the IEEE
Conference_Location :
New Orleans, LA, USA
Print_ISBN :
0-7803-0785-2
DOI :
10.1109/IEMBS.1988.94599