• DocumentCode
    3428455
  • Title

    A bioimplantable telemetered blood pressure measurement system for transgenic mice

  • Author

    Maymandi-Nejad, Mohammad ; Sachdev, Manoj

  • Author_Institution
    Waterloo Univ., Ont., Canada
  • fYear
    2004
  • fDate
    1-3 Dec. 2004
  • Abstract
    A system including an implantable integrated circuit (IC) for measuring the blood pressure of transgenic mice is implemented in 0.18 μm CMOS technology. The system specifications and circuit design challenges are summarized and discussed in this paper. The implant part of the system is to operate with a single miniature battery for more than two month. The measured data is transmitted wirelessly to a base station. The power consumption of the implantable system is 38 μW.
  • Keywords
    CMOS integrated circuits; biomedical telemetry; blood pressure measurement; network synthesis; prosthetics; 38 muW; CMOS technology; bioimplantable telemetered blood pressure measurement system; circuit design; implantable integrated circuit; single miniature battery; transgenic mice; Battery charge measurement; Blood pressure; CMOS integrated circuits; CMOS technology; Circuit synthesis; Implants; Integrated circuit measurements; Integrated circuit technology; Mice; Pressure measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Circuits and Systems, 2004 IEEE International Workshop on
  • Print_ISBN
    0-7803-8665-5
  • Type

    conf

  • DOI
    10.1109/BIOCAS.2004.1454191
  • Filename