  • Abstract
    A double-balance bridge and a measurement set-up based on the three-voltmeter method are used to calibrate inductance standards by linking them with reference capacitors or resistors. For inductances from 1 mH to 100 mH and frequencies up to 5 kHz, the calibration uncertainties are less than 1 part in 104
  • Keywords
    bridge instruments; calibration; inductance measurement; measurement standards; measurement uncertainty; voltmeters; calibration uncertainties; double-balance bridge; inductance measurement; inductance standards; measurement standards; three-voltmeter method; Bridge circuits; Calibration; Capacitors; Inductance measurement; Joining processes; Measurement standards; Phase transformers; Resistors; Transformer cores; Voltage;