DocumentCode
3445050
Title
Stainless steel particulate and human tissue responses
Author
Perry, Wes L. ; Lemons, Jack E. ; Meyer, Richard D. ; Clark, Laura
Author_Institution
Dept. of Surg., Alabama Univ., Birmingham, AL, USA
fYear
1997
fDate
4-6 Apr 1997
Firstpage
283
Lastpage
284
Abstract
Biomaterial particulate properties, such as chemistry, size, shape, and number, are key factors in determining biological performance of implants. Debris has been shown to elicit inflammatory responses and possibly lead to osteolysis. One objective of this study is to provide a more in-depth understanding of human tissue responses to 316L stainless steel particulates. The histological interactions and overall metallic contents of tissues collected from cross-screw stabilized, explanted intramedullary rods were qualitatively analyzed via light microscopy. More metallic debris was detected at sites near locking screws, where metal-on-metal motion existed. However, no direct correlations were observed between debris amounts and inflammatory response scores. There was also no direct correlation observed between inflammatory response scores in tissues near locking screws when compared to tissues away from sites of mechanical motion and debris generation. The results of this study represent responses over a relatively short period of time, typically less than three years, as the IM nails were utilized as temporary fixation devices
Keywords
austenitic stainless steel; biomechanics; cellular biophysics; optical microscopy; particle size; prosthetics; 316L stainless steel particulates; IM nails; biological performance; biomaterial particulate properties; chemistry; cross-screw stabilized explanted intramedullary rods; debris generation; histological interactions; human tissue responses; implants; inflammatory responses; light microscopy; locking screws; mechanical motion; metal-on-metal motion; metallic contents; metallic debris; number; osteolysis; shape; size; stainless steel particulate; temporary fixation devices; Biological materials; Chemistry; Fasteners; Humans; Implants; Microscopy; Motion detection; Orthopedic surgery; Shape; Steel;
fLanguage
English
Publisher
ieee
Conference_Titel
Biomedical Engineering Conference, 1997., Proceedings of the 1997 Sixteenth Southern
Conference_Location
Biloxi, MS
ISSN
1086-4105
Print_ISBN
0-7803-3869-3
Type
conf
DOI
10.1109/SBEC.1997.583286
Filename
583286
Link To Document