Title :
Determination of DC Voltage Ratio of Self-Calibrating Divider
Author :
Liu, L.X. ; Chua, S.W. ; Ang, C.K.
Author_Institution :
National Metrol. Centre of Singapore Stand.
Abstract :
This paper presents two techniques to determine the ratios of resistive dc voltage dividers up to 1 kV. The techniques are based on absolute calibration methods. The set-up requires equipment commonly available in most standard laboratory and is capable of providing high accuracy calibration under practical laboratory working conditions
Keywords :
calibration; voltage dividers; voltage measurement; 1 kV; DC voltage dividers; DC voltage ratio; absolute calibration; self-calibrating divider; Calibration; Circuit testing; Detectors; Employee welfare; Laboratories; Metrology; Productivity; Switches; Technological innovation; Voltage;
Conference_Titel :
Precision Electromagnetic Measurements Digest, 2004 Conference on
Conference_Location :
London
Print_ISBN :
0-7803-8494-6
Electronic_ISBN :
0-7803-8494-6
DOI :
10.1109/CPEM.2004.305517