Title :
Comparing Different Methods for the Calibration of the mV Ranges of an AC Measurement Standard
Author_Institution :
Fluke Corp., Everett, VA
Abstract :
A new method has been developed for the calibration of an AC measurement standard at 20 mV and below using a resistive voltage divider (RVD). Measurements made with the RVD are compared to measurements using the existing "bootstrap" method given in the service manual and also to measurements using an AC/DC transfer standard
Keywords :
calibration; electric current measurement; measurement standards; voltage dividers; AC measurement standard; AC/DC transfer standard; resistive voltage divider; Calibration; Frequency conversion; Frequency measurement; Inductors; Laboratories; Low-frequency noise; Measurement standards; Noise cancellation; Testing; Voltage;
Conference_Titel :
Precision Electromagnetic Measurements Digest, 2004 Conference on
Conference_Location :
London
Print_ISBN :
0-7803-8494-6
Electronic_ISBN :
0-7803-8494-6
DOI :
10.1109/CPEM.2004.305521