DocumentCode
3447091
Title
DC Resistive Divider for 10 V HTS Voltage Calibrator
Author
Klushin, A.M. ; Katkov, A.S. ; Pesel, E. ; Warnecke, P.
Author_Institution
Inst. fur Schichten und Grenzflachen, Forschungszcntrum Julich GmbH
fYear
2004
fDate
38139
Firstpage
332
Lastpage
333
Abstract
Resistive divider for a HTS voltage calibrator (Klushin, 2003) with an output voltage up to 10 V was investigated. The developed self-calibration method and parameters of the main parts of the divider showed, that a combined uncertainty of the voltage ratios in the range from 20 mV up to 10 V does not exceed a few parts in 108
Keywords
calibration; voltage dividers; voltage measurement; 10 V; DC resistive divider; HTS voltage calibrator; self-calibration method; uncertainty; Calibration; Diodes; Electrical resistance measurement; High temperature superconductors; Metrology; Resistors; Stability; Thermal force; Uncertainty; Voltage;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements Digest, 2004 Conference on
Conference_Location
London
Print_ISBN
0-7803-8494-6
Electronic_ISBN
0-7803-8494-6
Type
conf
DOI
10.1109/CPEM.2004.305600
Filename
4097255
Link To Document