• DocumentCode
    3447214
  • Title

    An Environment Stabilized Super High Resistance and Its Measurement

  • Author

    Haiming, Shao ; Feipeng, Lin ; Bo, Liang ; Hao, Wang ; Li Zhibin

  • Author_Institution
    Nat. Inst. of Metrol., Beijing
  • fYear
    2004
  • fDate
    38139
  • Firstpage
    344
  • Lastpage
    345
  • Abstract
    The development of DC high resistance standards from 10 MOmega to 100 TOmega and a dual-voltage-sources bridge is described. The resistance standards are significantly less sensitive to humidity and temperature variations as they are mechanically sealed in a thermostat chamber. A method, for calibrating 10 MOmega to 100 TOmega resistors, is described with results and uncertainty budgets reported
  • Keywords
    calibration; electric resistance measurement; measurement standards; measurement uncertainty; resistors; 10 to 100E6 Mohm; DC high resistance standards; calibration method; dual-voltage-sources bridge; mechanically sealed; thermostat chamber; uncertainty budgets; Bridge circuits; Connectors; Current measurement; Electrical resistance measurement; Humidity; Metrology; Resistors; Standards development; Temperature sensors; Voltage; Dual-Voltage-Sources Bridge; High Resistance;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 2004 Conference on
  • Conference_Location
    London
  • Print_ISBN
    0-7803-8494-6
  • Electronic_ISBN
    0-7803-8494-6
  • Type

    conf

  • DOI
    10.1109/CPEM.2004.305606
  • Filename
    4097261