• DocumentCode
    3447529
  • Title

    Evaluation of the Frequency Dependence of the Resistance and Capacitance Standards in the BIPM Quadrature Bridge

  • Author

    Delahaye, F. ; Goebel, R.

  • Author_Institution
    Bur. Int. des Poids et Mesures, Sevres
  • fYear
    2004
  • fDate
    38139
  • Firstpage
    372
  • Lastpage
    373
  • Abstract
    The BIPM uses a quadrature bridge to link capacitance standards to resistance standards. The frequency dependence of the resistances used in the quadrature bridge (a pair of 51.6 kOmega resistors) is determined by comparison with a calculable coaxial resistor of 1290.6 Omega. The quadrature bridge can be operated at three different frequencies (1027 Hz, 1541 Hz and 3082 Hz) using three different pairs of capacitors of values 3000 pF, 2000 pF and 1000 pF, respectively. The frequency dependence of each capacitance in this range of frequencies is deduced from these measurements
  • Keywords
    capacitance measurement; electric resistance measurement; measurement standards; resistors; 1027 Hz; 12.90 ohm; 1541 Hz; 3082 Hz; 51.6 kohm; BIPM; capacitance standards; coaxial resistor; frequency dependence; quadrature bridge; resistance standards; Bridge circuits; Calibration; Capacitance measurement; Capacitors; Coaxial components; Electrical resistance measurement; Frequency dependence; Frequency measurement; Measurement standards; Resistors;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 2004 Conference on
  • Conference_Location
    London
  • Print_ISBN
    0-7803-8494-6
  • Electronic_ISBN
    0-7803-8494-6
  • Type

    conf

  • DOI
    10.1109/CPEM.2004.305620
  • Filename
    4097275