Abstract
We describe a novel cylindrical TE01delta mode dielectric resonator used to measure the complex permittivity of liquids. We performed measurements at frequencies of 422 MHz and 970 MHz on 1-propanol alcohol, methyl alcohol, deionized water, and conductivity-standard solutions of water and potassium chloride. Rigorous mode-matching analysis was used to determine the complex permittivity. Type B measurement uncertainties were estimated to be less than 0.01 for the relative permittivity and less than 0.01 for the loss tangent
Keywords
dielectric resonators; measurement uncertainty; permittivity measurement; 1-propanol alcohol; 422 MHz; 970 MHz; cylindrical mode dielectric resonator; deionized water; measurement uncertainty; methyl alcohol; mode-matching analysis; potassium chloride; relative permittivity measurement; Dielectric liquids; Dielectric loss measurement; Dielectric measurements; Frequency measurement; Loss measurement; Magnetic field measurement; Measurement uncertainty; Permittivity measurement; Sea measurements; Tellurium;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements Digest, 2004 Conference on
Conference_Location
London
Print_ISBN
0-7803-8493-8
Electronic_ISBN
0-7803-8494-6
Type
conf
DOI
10.1109/CPEM.2004.305314
Filename
4097325
Link To Document