• DocumentCode
    3448748
  • Title

    Improved Capacitance Measurements with Respect to a 1 pf Cross Capacitor from 200 Hz to 2000 Hz

  • Author

    Wang, Yicheng ; Shields, Scott

  • Author_Institution
    Quantum Electr. Metrol. Div., Nat. Inst. of Stand. & Technol., Gaithersburg, MD
  • fYear
    2004
  • fDate
    38139
  • Firstpage
    489
  • Lastpage
    490
  • Abstract
    We describe frequency dependence measurements of fused-silica capacitance standards from 200 Hz to 200 Hz, using a 1 pF cross capacitor as the reference. The relative combined standard uncertainty at 200 Hz (the largest in the frequency range) is 0.11times 10-6 , which is smaller by about a factor of two than the uncertainty reported earlier from NIST
  • Keywords
    capacitance measurement; capacitors; measurement standards; 1 pF; 200 to 2000 Hz; capacitance measurements; capacitance standards; cross capacitor; frequency dependence measurements; standard uncertainty; Bridge circuits; Calibration; Capacitance measurement; Capacitors; Electric variables measurement; Electrodes; Frequency dependence; Frequency measurement; NIST; Quantum capacitance;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 2004 Conference on
  • Conference_Location
    London
  • Print_ISBN
    0-7803-8494-6
  • Electronic_ISBN
    0-7803-8494-6
  • Type

    conf

  • DOI
    10.1109/CPEM.2004.305325
  • Filename
    4097336