Title :
Inductance Calibration Method Using a Commercial Lcr Meter
Author :
Yonenaga, A. ; Nakamura, Y.
Author_Institution :
National Metrol. Inst. of Japan, National Inst. of Adv. Ind. Sci. & Technol., Tsukuba
Abstract :
A simple calibration method, which determines inductance from the capacitance standard using a commercially available LCR meter, is presented. Two 10 mH standard inductors were calibrated against a 1 muF reference capacitor at 1592 Hz. The relative combined standard uncertainty was estimated to be 7.4 muH/H
Keywords :
calibration; inductance measurement; inductors; measurement standards; measurement uncertainty; meters; 1 muF; 1592 Hz; LCR meter; capacitance standard; inductance calibration method; measurement standards; measurement uncertainty; standard uncertainty; Bridge circuits; Calibration; Capacitance; Capacitors; Frequency; Impedance; Inductance; Inductors; Metrology; Standards development;
Conference_Titel :
Precision Electromagnetic Measurements Digest, 2004 Conference on
Conference_Location :
London
Print_ISBN :
0-7803-8494-6
Electronic_ISBN :
0-7803-8494-6
DOI :
10.1109/CPEM.2004.305397