Multivalue High Voltage Loss Angle Standard
Author
Klionsky, M.D. ; Stepenkov, V.V.
Author_Institution
D. I. Mendeleyev Inst. for Metrol., St. Petersburg
fYear
2004
fDate
38139
Firstpage
606
Lastpage
607
Abstract
The standard of the loss angle tangent (D) is intended for calibration of capacitance bridges at the voltage 1-10 kV and frequency 50 Hz. A high repeatability of the D values is obtained due to special design, and the usage of small-size ceramic capacitors makes the standard portable. The expanded uncertainty of the output parameters of the standard is (1-2)middot10-5+0,001D for the loss angle tangent and 0,1 % (k=2) for the capacitance
Keywords
bridge instruments; calibration; ceramic capacitors; dielectric loss measurement; measurement standards; measurement uncertainty; 1 to 10 kV; 50 Hz; bridge instruments; capacitance bridge calibration; ceramic capacitors; loss angle tangent; measurement standards; multivalue high voltage loss angle standard; Bridge circuits; Calibration; Capacitance measurement; Capacitors; Ceramics; Frequency; Lifting equipment; Metrology; Resistors; Voltage;