  • Abstract
    In this work we report a simple time of flight technique for assessing the number and motility of pig sperm in a semen sample. Acoustic plate mode (APM) devices have been employed as the sensor element at the end of a channel down which sperm swim. Combining the APM data with conventional microscopy, an estimate of 18.9 plusmn4.8 pg for the pig sperm effective mass at 52 MHz APM devices is derived.
  • Keywords
    acoustic devices; biomedical equipment; biomedical measurement; cell motility; acoustic plate mode devices; frequency 52 MHz; microscopy; sensor element; sperm effective mass; sperm motility assessment; time-of-flight technique; Acoustic devices; Artificial intelligence; Biomedical acoustics; Coatings; Face detection; Fingers; Frequency; Instruments; Low pass filters; Time measurement;