• DocumentCode
    3452473
  • Title

    High-Precision Ventilator with Compensation Algorithm and Automatic Regulation

  • Author

    An, Peng ; Song, Jia-Tao

  • Author_Institution
    Sch. of Electron. & Inf. Eng., Ningbo Univ. of Technol., Ningbo, China
  • fYear
    2010
  • fDate
    27-28 Nov. 2010
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    CPAP (Continuous Positive Airway Pressure) ventilator is designed based on compensation algorithm and Automatic Regulation. With the use of patient status and system changes in the environment, breathing machine through the compensation algorithm will be dynamically adjusted accordingly. The compensation algorithm, taking into account regional differences in altitude, the pressure, masks and flow loop caused by the pressure of the gas leak and pressure conditioned lack of progressive features. MPC5554 is chosen as the main processor in the system for its high integration and precision. It has several interfaces to communicate with external devices. In the process of designing, some modules are customized. Compared with similar products, the volume and weight are decreased and the cost is reduced.
  • Keywords
    biomedical engineering; compensation; medical computing; microcontrollers; patient care; patient diagnosis; patient monitoring; patient treatment; pressure; regulation; ventilation; MPC5554; automatic regulation; breathing machine; compensation algorithm; continuous positive airway pressure ventilator; gas leak pressure; high precision ventilator; Algorithm design and analysis; DC motors; Heuristic algorithms; Lungs; Pulse width modulation; Radiation detectors; Sleep apnea;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Database Technology and Applications (DBTA), 2010 2nd International Workshop on
  • Conference_Location
    Wuhan
  • Print_ISBN
    978-1-4244-6975-8
  • Electronic_ISBN
    978-1-4244-6977-2
  • Type

    conf

  • DOI
    10.1109/DBTA.2010.5658990
  • Filename
    5658990