DocumentCode
3455137
Title
Development of a multichannel ultrasonic interferometer for the measurements under high pressures and its application to biomolecular studies
Author
Sarvazyan, A.P. ; Chalikian, T.V.
Author_Institution
Inst. of Biol. Phys., Acad. of Sci., Pushchino, USSR
fYear
1988
fDate
2-5 Oct 1988
Firstpage
937
Abstract
A description is given of a device that comprises four parallel fixed-path interferometric cells with a volume of 0.2 cc each. This device provides relative measurements of sound velocity with an error of about 10-3 to 10-4% in the pressure range of 1 to 2500 bar. The specific increments of the nonlinearity acoustic parameter B /A , for different biomolecular compounds (nucleotides, amino acids, and proteins), have been measured. The error of relative measurements of B /A is less than 0.5%
Keywords
biological techniques and instruments; high-pressure techniques; molecular biophysics; ultrasonic equipment; ultrasonic measurement; 1 to 2500 bar; amino acids; biomolecular studies; high pressures; multichannel ultrasonic interferometer; nonlinearity acoustic parameter; nucleotides; proteins; sound velocity; Acoustic applications; Acoustic measurements; Equations; Nonlinear acoustics; Solvents; Sugar; Temperature; Thermal expansion; Thermodynamics; Ultrasonic variables measurement;