• DocumentCode
    3455137
  • Title

    Development of a multichannel ultrasonic interferometer for the measurements under high pressures and its application to biomolecular studies

  • Author

    Sarvazyan, A.P. ; Chalikian, T.V.

  • Author_Institution
    Inst. of Biol. Phys., Acad. of Sci., Pushchino, USSR
  • fYear
    1988
  • fDate
    2-5 Oct 1988
  • Firstpage
    937
  • Abstract
    A description is given of a device that comprises four parallel fixed-path interferometric cells with a volume of 0.2 cc each. This device provides relative measurements of sound velocity with an error of about 10-3 to 10-4% in the pressure range of 1 to 2500 bar. The specific increments of the nonlinearity acoustic parameter B/A, for different biomolecular compounds (nucleotides, amino acids, and proteins), have been measured. The error of relative measurements of B/A is less than 0.5%
  • Keywords
    biological techniques and instruments; high-pressure techniques; molecular biophysics; ultrasonic equipment; ultrasonic measurement; 1 to 2500 bar; amino acids; biomolecular studies; high pressures; multichannel ultrasonic interferometer; nonlinearity acoustic parameter; nucleotides; proteins; sound velocity; Acoustic applications; Acoustic measurements; Equations; Nonlinear acoustics; Solvents; Sugar; Temperature; Thermal expansion; Thermodynamics; Ultrasonic variables measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 1988. Proceedings., IEEE 1988
  • Conference_Location
    Chicago, IL
  • Type

    conf

  • DOI
    10.1109/ULTSYM.1988.49514
  • Filename
    49514