    Blood pressure measurement device based on the arterial tonometry method with micro triaxial force sensor

  • Author

    Shioya, K. ; Dohi, Tadashi

  • Author_Institution
    Chuo Univ., Tokyo, Japan
  • fYear
    2013
  • fDate
    16-20 June 2013
  • Firstpage
    2389
  • Lastpage
    2392
  • Abstract
    This paper reports on a blood pressure measurement device for measuring blood pressure by the arterial tonometry method with micro triaxial force sensor. The micro triaxial force sensor consists of two pressure sensor elements and six shear stress sensor elements. These sensor elements can measure the vertical and horizontal components of blood pressure. Additionally, the four shear stress sensor elements distinguish verticality pressurization and tilted pressurization. Therefore the accuracy of the blood pressure was improved. Errors of blood pressure were 5.4 % at maximum blood pressure and 7.9 % at minimum blood pressure. The error of the fabricated device was smaller than the device with only pressure sensor.
  • Keywords
    blood pressure measurement; force sensors; pressure sensors; arterial tonometry; blood pressure measurement device; micro triaxial force sensor; pressure sensor; shear stress sensor; tilted pressurization; verticality pressurization; Arteries; Blood pressure; Force; Force measurement; Force sensors; Pressure measurement; Stress; Arterial Tonometry method; Blood pressure measurement; Micro triaxial force sensor;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Solid-State Sensors, Actuators and Microsystems (TRANSDUCERS & EUROSENSORS XXVII), 2013 Transducers & Eurosensors XXVII: The 17th International Conference on
  • Conference_Location
    Barcelona
  • Type

    conf

  • DOI
    10.1109/Transducers.2013.6627287
  • Filename
    6627287