DocumentCode
3463900
Title
Standard measuring system for calibration of magnetic flux density gradiometers
Author
Khorev, V.N. ; Shifrin, Ya V. ; Shubin, S.A. ; Park, P.G.
Author_Institution
D.I. Mendeleyev Inst. for Metrol. (VNIIM), St. Petersburg, Russia
fYear
2010
fDate
13-18 June 2010
Firstpage
314
Lastpage
315
Abstract
A standard measuring system (setup) has been developed at VNIIM to be used for the calibration of magnetic flux density gradient (MFDG) coils, sensors and magnetic gradiometers (MG). This measurement standard is based on the calculable quartz coil used for the reproduction of gradient which is uniform in a specified working space. It also includes measuring equipment to compare coils, gradiometers and differential measuring transducers. Theoretical and experimental aspects of the problem are discussed. The developed standard system allows calibrating MFDG coils with an uncertainty of 0.1 % within the range from 0.005 to 0.5 T/m.
Keywords
calibration; fluxgate magnetometers; magnetic flux; measurement standards; MFDG coils; VNIIM; differential measuring transducers; magnetic flux density gradient; magnetic gradiometers calibration; measurement standard; measuring equipment; quartz coil; sensors; standard measuring system; Calibration; Coils; Density measurement; Extraterrestrial measurements; Magnetic flux; Magnetic flux density; Magnetic sensors; Measurement standards; Sensor systems; Standards development;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements (CPEM), 2010 Conference on
Conference_Location
Daejeon
Print_ISBN
978-1-4244-6795-2
Type
conf
DOI
10.1109/CPEM.2010.5543530
Filename
5543530
Link To Document